VAYU Breathology · Project scope and technology roadmap

From one breath platform to a portfolio of applications, funded one proven result at a time.

The G-300 is the team's base technology, with regulatory status confirmed in Thailand and Singapore. New funding extends it to further applications, and each one must prove its breath signal before larger spending is released.

TRL 8
G-300 platform, within its confirmed Thai and Singapore scope
6 programs
New applications, each with its own stop rule
Year 3
Nearest milestone: ketosis tracking at TRL 8
US: none
No medical device status in the United States today

Where we stand

Regulatory status differs by market

Everything in this roadmap is read against these three positions.

Bangkok skyline
Status confirmed

Thailand

Thai FDA status confirmed for the G-300. Registration details are available to investors on request.

Singapore skyline across Marina Bay
Status confirmed

Singapore

Singapore regulatory status confirmed for the G-300. Registration details are available to investors on request.

San Diego skyline
No medical device status

United States

VAYU has no FDA medical device status. The US product is positioned for general wellness only, and nothing on this page applies to it as a medical claim.

Scientist holding up a test tube in a laboratory

First clinical evidence, from one Thai hospital

71participants
ICC > 0.9breath vs venous glucose
1.23 mg/dLmean difference

A single-site, cross-sectional study of the G-300 in Thailand. Its authors call for further studies to confirm accuracy and clinical use, so replication is the first use of funds. It does not apply to the US wellness product. Read the study

Project scope

What this funding covers, and what it does not

The project reuses the G-300 device, cartridge format, app and regulatory experience. That shortens the engineering. It does not shorten the biology, so each new application starts by testing whether its signal exists.

Engineers planning at a whiteboard
WORKSTREAM 1

Evidence base

Multi-site glucose replication, reporting MARD and Clarke error grid.

WORKSTREAM 2

New applications

Ketosis tracking first, then kidney, heart failure, liver, Parkinson's and Alzheimer's.

WORKSTREAM 3

Sensor platform

New cartridge channels, plus a next-generation LIG sensor and imprinted coatings with an adopt-or-drop decision.

WORKSTREAM 4

Regulatory

New-indication filings in Thailand and Singapore, plus a US route assessment.

WORKSTREAM 5

Quality and data

ISO 13485 and risk management, ethics approvals, and a reference breath data set.

In scope

  • Studies against a reference instrument (GC-MS or SIFT-MS) and clinical standards
  • Device, cartridge and software changes needed for new applications
  • Partner-led studies for Parkinson's and Alzheimer's
  • An FDA pre-submission once multi-site data exists

Out of scope

  • Any medical, disease or glucose claim for the US product before FDA authorization
  • Advice on medication or treatment, including diuretic dosing
  • Replacing blood glucose testing in any market without a study that supports it
  • LIG or imprinting performance claims before testing at breath humidity

Next-generation sensor research

How LIG, functionalization and molecular imprinting work

This section describes the research track in Workstream 3. It makes no claim about the current G-300 sensor, and none of these coatings is part of a cleared product.

Laser-induced graphene

A CO₂ laser scans a polyimide film in ordinary air. Local temperatures above 2,500 °C break the polymer's bonds, oxygen and nitrogen leave as gases, and the remaining carbon rearranges into graphene. The escaping gases leave a conductive 3D foam.

The attraction for sensing is a large, porous surface that can be patterned directly into electrodes, with no separate graphene transfer step. Bare LIG lacks selectivity for targets such as acetone, so it needs a coating.

Laser
10.6 µm CO₂
Surface area
~340 m²/g
Sheet resistance
<15–35 Ω/sq
Carbon content
71% → ~97%
1. Start with polyimide filmPolyimide2. Laser writes porous graphenePolyimide3. Coat it and read resistancePolyimideCO₂ laser, in airGases leavePorous grapheneResistanceCoatingBreath gas binds to coating
Figures from Lin et al. 2014 and vary with laser settings. In a chemiresistor, the target gas changes the material's charge carriers, and the device reads this as a change in resistance.

Functionalization: giving each channel a job

Functionalization means adding a material to the graphene surface so a channel responds more strongly to its target gas. Each approach works differently and has reached a different stage.

ApproachTargetHow it worksPublished resultStatus
Polyaniline (PANI)AmmoniaAmmonia takes protons from the polymer, which removes charge carriers and raises resistance. Returning to clean air reverses this.Tested at 5 ppm and above, with an 18-minute response. The 2.38 ppb limit is calculated. Response fell 30% by day 45.Not yet at breath levels
Metal oxides (such as ZnO)AcetoneA junction forms where the oxide meets graphene. Gas landing on the surface changes how charge crosses that junction, and this improves selectivity over LIG alone.−24% at 1 ppm, with a 21-second response. A 4 ppb limit is reported.Closest to breath levels
Noble metals (Pt, Pd)To be determinedThe metal splits gas molecules, and the fragments move onto the support (spillover), which strengthens the electrical change.Shown mainly for hydrogen. No LIG result for breath VOCs was found.Unproven for breath
Water barriers (PDMS, molecular sieve)All channelsA layer over the sensor slows or blocks water while letting the target gas through.PDMS held NOx response steady up to 90% humidity, but response took about 15 minutes at 25 °C.Works, with a speed cost
Sensor arraysMixturesEach channel responds differently, and software reads the overall pattern as a fingerprint.MXene on LIG electrodes scored 88.9% in a blind drink/no-drink test on volunteers.No disease study yet

Molecular imprinting: a cavity shaped for one molecule

A molecularly imprinted polymer (MIP) is cast around the target molecule. Removing the target leaves cavities that match its size, shape and chemical groups, so it should bind more readily than other gases. Because most MIPs conduct poorly, they are blended with conductive material or grown directly on an electrode. When the target binds, the polymer can swell and push conductive paths apart, which changes resistance.

1. Mix target and monomersLIG electrode2. Polymerize around itLIG electrode3. Remove the targetLIG electrode4. Target rebindsLIG electrodeOther gas blocked
The recommended starting point in gas-sensing work is about 4 monomers per template molecule. Incomplete removal leaves the target blocking its own binding sites. The diagram shows the ideal case, and real polymers are less selective.

What the evidence shows

  • LIG with an imprinted polymer has detected tetracycline at 0.32 nM, but in liquid, not in gas.
  • Conductive imprinted polymers have sensed acetophenone vapor, with a 42 ppm limit. Breath targets sit far lower, at parts per billion.
  • One acetone MIP responded about 10 times more strongly to acetone than to formaldehyde at 5 ppm.

Why breath is hard

  • Small, volatile molecules offer few binding points. One acetone MIP showed no selectivity over similar alcohols and ketones.
  • Gas-phase MIPs tend to show more non-specific binding than liquid-phase ones.
  • Humidity remains a challenge, and binding sites can degrade over time.
  • We found no published LIG and MIP sensor for breath or other gases.
How we will test it. Imprinting is an experimental option within the LIG track, starting with acetone. Each imprinted channel is compared with an otherwise identical non-imprinted control, at 90%+ humidity, in breath-like gas mixtures and over 30–90 days. It goes forward only if it clearly beats the control and the best conventional coating, with the decision made at the year 2.5 LIG gate.
Technical sources

Technology readiness

A mature platform, with most applications still early

Management assessment on the 1–9 TRL scale, where TRL 8 means production-equivalent units that pass final testing and are cleared for the claim made.

ReachedPossibly reachedFunded path to TRL 8

Ketosis at TRL 5 assumes the G-300 cartridge can report acetone. Kidney, liver and Alzheimer's ratings are provisional until a fuller evidence review is complete.

Roadmap

The path to TRL 8

Years from funding close. Programs continue only after passing their go/no-go gate, marked with a diamond. Durations are planning estimates.

Tranche 1Tranche 2Tranche 3CloseYr 1Yr 2Yr 3Yr 4Yr 5Yr 6Yr 7Yr 8Yr 9Yr 10PlatformGlucose evidenceGlucose evidence: Multi-site study, year 0 to 1.5Yr 1.5Next-gen LIG sensorNext-gen LIG sensor: Sensor channel and bench tests, year 0 to 1.5Next-gen LIG sensor: Handheld study, year 1.5 to 2.5Next-gen LIG sensor: Multi-site study, year 2.5 to 4Next-gen LIG sensor: Pivotal study and filing, year 4 to 5Next-gen LIG sensor: go/no-go gate at year 2.5Yr 5ApplicationsKetosis trackingKetosis tracking: Handheld study, year 0 to 1Ketosis tracking: Multi-site study, year 1 to 2Ketosis tracking: Pivotal study and filing, year 2 to 3Ketosis tracking: go/no-go gate at year 1Yr 3KidneyKidney: Signal check, year 0 to 1.5Kidney: Sensor channel and bench tests, year 1.5 to 2.5Kidney: Handheld study, year 2.5 to 3.5Kidney: Multi-site study, year 3.5 to 5Kidney: Pivotal study and filing, year 5 to 6.5Kidney: go/no-go gate at year 1.5Yr 6.5Heart failureHeart failure: Signal check, year 0 to 2Heart failure: Sensor channel and bench tests, year 2 to 3Heart failure: Handheld study, year 3 to 4Heart failure: Multi-site study, year 4 to 5.5Heart failure: Pivotal study and filing, year 5.5 to 7Heart failure: go/no-go gate at year 2Yr 7LiverLiver: Signal check, year 1 to 2.5Liver: Sensor channel and bench tests, year 2.5 to 3.5Liver: Handheld study, year 3.5 to 4.5Liver: Multi-site study, year 4.5 to 6Liver: Pivotal study and filing, year 6 to 7.5Liver: go/no-go gate at year 2.5Yr 7.5Parkinson'sParkinson's: Signal check, year 0 to 2.5Parkinson's: Sensor channel and bench tests, year 2.5 to 4Parkinson's: Handheld study, year 4 to 5.5Parkinson's: Multi-site study, year 5.5 to 7.5Parkinson's: Pivotal study and filing, year 7.5 to 9.5Parkinson's: go/no-go gate at year 2.5Yr 9.5Alzheimer'sAlzheimer's: Signal check, year 0 to 3Alzheimer's: Sensor channel and bench tests, year 3 to 4.5Alzheimer's: Handheld study, year 4.5 to 6Alzheimer's: Multi-site study, year 6 to 8Alzheimer's: Pivotal study and filing, year 8 to 10Alzheimer's: go/no-go gate at year 3Yr 10+US optionUS FDA routeUS FDA route: FDA pre-submission, year 1.5 to 3US FDA route: go/no-go gate at year 3Decision
Signal checkSensor channel and bench testsHandheld studyMulti-site studyPivotal study and filingFDA pre-submissionGo/no-go gate

What each program must prove

ProgramTRL todayReaches TRL 8First proof pointCompared againstStop if
Ketosis tracking5About year 3Follows personal ketone trends in 30–50 peopleBlood ketonesIt cannot follow trends in people outside model training
Kidney2–3About year 6.5Breath ammonia tracks kidney markers, starting with dialysis patientsBUN, creatinineThe signal mostly reflects the mouth, not the blood
Heart failure2–3About year 7Breath changes before decompensation in patients followed at homeClinical eventsChanges do not come early enough to act on
Liver2About year 7.5A marker predicts steatosis apart from diet and weightMRI-PDFF, FibroScanNo independent marker is found
Parkinson's3About year 9.5Named compounds replicate at a second siteClinical diagnosis, DaTscanCompounds do not replicate
Alzheimer's2–3Year 10 or laterBreath adds information beyond a blood testPlasma p-Tau, amyloid PETBreath adds nothing to a blood test

Disease programs begin with a reference instrument alongside the G-300, which tests the breath signal before any new sensor channel is built.

United States

The US stays wellness-only until the FDA says otherwise

A US medical route is an option this roadmap may pursue. It is not a commitment, and no timeline for FDA authorization is promised.

Current US status: no medical device status. US product, app and marketing carry no disease scores, glucose readings or medical claims.
  1. NowGeneral wellness positioning only, reviewed by an FDA regulatory consultant and kept separate from clinical work elsewhere.
  2. Year 0–1.5Build multi-site evidence in Thailand and Singapore, using protocols designed so the data could support a later FDA discussion.
  3. Year 1.5–3Pick one indication and hold an FDA pre-submission meeting to test the likely route, such as De Novo or 510(k).
  4. Year 3 gateProceed only if the FDA feedback, the data and the cost of a US study support it. Otherwise the US stays wellness-only.

Funding

Funding tied to results

Each tranche is released on evidence, so investors can see what the previous stage proved before committing more. Tranche amounts are shared directly with investors.

Close-up of microchips on a circuit board

Tranche 1: prove it (0–18 months)

  • Replicate glucose accuracy across several sites
  • Take ketosis tracking to TRL 6
  • Complete the kidney signal check
  • Enroll the heart failure home-monitoring cohort
  • Test the LIG sensor at breath humidity
  • Start partner-led Parkinson's and Alzheimer's signal checks, then liver from month 12

Release gate: glucose replicated, ketosis at TRL 6, kidney decision made.

Tranche 2: narrow and build (18–36 months)

  • Take ketosis tracking to TRL 8 and file
  • Make the heart failure go/no-go decision
  • Build sensor channels for programs that passed
  • Decide whether to adopt or drop LIG
  • Read out liver and Parkinson's signal checks at month 30
  • Hold the FDA pre-submission if the data supports it

Release gate: ketosis filed, and at least one disease program past its gate.

Tranche 3: scale winners (36 months onward)

  • Run multi-site and pivotal studies for surviving programs
  • Prepare new-indication filings, starting in Thailand and Singapore
  • Decide the US FDA route and assess wider ASEAN markets
  • Seek partner funding for Parkinson's and Alzheimer's

Outcome: one or more new regulated applications on the G-300 platform.

Risks

Main technical risks

Scientific

  • The glucose result may not replicate. Earlier breath studies vary widely, so replication comes first.
  • Disease signals may be weak. Published work is mostly small pilots, so gates limit spending.
  • Diet, medication and age may confound results. Protocols control for them, and models are tested on unseen people.

Engineering and regulatory

  • New targets may need new channels. Pentane and ammonia are hard to measure at breath levels.
  • Humidity and drift limit next-generation sensors. LIG stays optional until it passes testing.
  • New claims may raise device class. Classification advice is sought before each pivotal study.

Talk to the team

Ask for registration details, the study data and tranche amounts.

[email protected]